April 2004
2-12 Basic Programming
Entering or Changing
Probe Configuration
Data (cont.)
Entering Calibration
Data
The Series 1 needs calibration data for moisture, oxygen, and
pressure probes, as well as auxiliary inputs. GE Panametrics has
already entered calibration data for probes; however, you should
verify this data before startup. You will need to enter this data if you:
sent the probes back to the factory for calibration
use a different probe
use a non-GE Panametrics input
To enter, verify, or change calibration data for any probe type, enter
the System Calibration Menu as described in Table 2-3 below. (see
Figure 3-19 on page 3-32 for a menu map.)
.
*The Settings Menu will prompt for a passcode (see page 2-6).
Table 2-2: Probe Types
Measurement Mode Probe Type
OXY (Oxygen)
(None)
% (Percent)
ppM (Parts per Million)
ppB (Parts per Billion)
HYG (Hygrometry)
(None)
MIS (MIS Probe)
TF (Three-Function Probe)
Mxx (M Series Probe)
Kh (Constant Dew Point)
T (Temperature)
(None)
MIS (MIS probe)
TF (Three-Function probe)
Mxx (M-Series probe)
Kt (Constant Temperature)
P (Pressure)
(None)
MIS (MIS probe)
TF (Three-Function probe)
Au1 (Auxiliary 1)
Au2 (Auxiliary 2)
Kp (Constant Pressure)
Aux1 (Auxiliary 1)
– (None)
I (Current)
V (Voltage)
Aux2 (Auxiliary 2)
– (None)
I (Current)
V (Voltage)
Table 2-3: Moving to the System Calibration Menu
Press the Main Menu key: To enter the:
SETTINGS* Settings Menu
SYSTEM System Settings Menu
CALIB System Calibration Menu