Basic Programming 2-11
April 2004
Entering or Changing
Probe Configuration
Data
Figure 3-19 on page 3-32 for a menu map.
1. From the Probe Configuration Menu (Table 2-1 on page 2-9),
move the pointer to the channel and probe you want to select. The
arrow will move only to installed channels.
2. Press [YES].
3. The possible probe types appear on the message line at the bottom
of the screen. Table 2-2 on page 2-12 shows the available probe
types.
If you choose to enter a constant for moisture (Kh), temperature (Kt),
or pressure (Kp), you must enter a constant value in the User Constant
Menu. See Entering Constants and User Functions on page 3-9 for
instructions.
4. Move the brackets to the correct probe type, then press [YES]. The
selected probe type displays.
5. Repeat Steps 1 through 4 for each channel.
6. To exit, press the DONE menu key until Main Menu appears on
the message line.
Note: You can only select probes for an installed channel.
If you are only using Moisture Image Series probes, proceed to
Displaying Measurements on page 2-28. If you are using any other
type of input device, proceed to Entering Calibration Data
on page 2-12.