Features and Capabilities 1-3
June 2003
BASEEFA Certification Rack, bench, panel Series 1s s/n 2000 and above and WPF Series 1s
s/n 2300 and above are BASEEFA certified to supply intrinsically
safe levels. Installation requires that the Series 1 be mounted in a
general purpose area only. M Series, TF Series, and Moisture Image
Series probes and Delta F oxygen cells are intrinsically safe when
connected to the Series 1, allowing for sensor installation in IEC/
CENELEC zone 0 areas.
CSA Certification Newer versions of rack, bench, panel and WPF Series 1’s are
CSA-NRTL certified intrinsically safe. Installation requires the
Series 1 be mounted in either a NEC CL.1, Div. 2, Group B, C or D
hazardous area or a general purpose area. M Series, TF Series, and
Moisture Image Series probes and Delta F oxygen cells are
intrinsically safe when connected to the Series 1, allowing for sensor
installation in NEC CL.1, Div. 1, Groups A, B, C and D hazardous
areas.
Note: Refer to certification labels on your instrument and
installation drawing #752-138 to determine the applicable
level of certification that your meter carries.