Advanced Programming 3-7
April 2004
Using Automatic
Calibration
The Series 1 automatically calibrates the moisture, pressure, and
oxygen measurement circuitry (Auto-Cal) at a user selected interval.
Auto-Cal compensates for any drift in the electronics. GE
Panametrics recommends setting the Auto-Cal interval to eight hours.
Set a smaller interval to Auto-Cal more frequently if the meter is
exposed to extreme temperatures or weather conditions.
Note: If you are using only Moisture Image Series Probes to
measure moisture, you do not have to Auto-Cal.
To set the Auto-Cal interval, press the keys shown in Table 3-3. The
screen appears similar to Figure 3-3 below. Refer to Figure 3-19 on
page 3-32 for a menu map.
* The Settings Menu will prompt for a passcode (see page 2-6).
Figure 3-3: Setting the Auto-Cal Interval
Table 3-3: Entering the Reference Menu
Press the Main Menu key: To enter the:
SETTINGS* Settings Menu
SERVICE System Service Menu
REFS Reference Menu
MOISTURE REFERENCE TABLE
Ch HIGH LOW
1
2
3
4
5
6
AutoCal Interval: 96 minutes.
Reference Menu
HELP INTERVAL PROBE DONE