Programming with PanaView 4-11
April 2004
Programming Error
Handling
1. To program the Series 1 for error handling, click on the “Errors”
tab. The screen appears similar to Figure 4-7 below.
2. Enter the desired values for the low and high error settings in the
appropriate text boxes.
3. To program the meter’s response to errors, pull down the “On Low
Error” list box, and click on one of five options:
• Ignore error
• Go to high limit
• Go to low limit
• Hold last value
• Go to error value.
4. Repeat this selection with the “On High Error” list box.
5. Click on [
Apply] to apply the changes immediately, and on [OK] to
confirm the changes and close the window. To return the window
to its previous settings, click on the [
Cancel] button.
Figure 4-7: Recorder Properties Window - Error Handling