2-20 Basic Programming
Entering Auxiliary Input
Calibration Data (cont.)
2. Use the CHANNEL menu key to cycle to the desired channel. The
channel number is indicated in the top right-hand corner of the
screen. The screen will only display installed channels.
3. Move the pointer to
ND and press [YES].
4. Enter the number of data points and press [YES]. The Series 1
requires two to 16 calibration data points. If the sensor is linear
you only need to enter two data points.
5. The pointer will move to LABEL. Press [YES] to enter the Label
Entry menu and enter or change the Label (as described in
“Making Tags” on page 2-38).
6. Move the pointer to the Units (or [Aux]) column heading. Press
[YES] to edit the Units label used when displaying the Auxiliary
Input (as described in “Making Tags” on page 2-38).
7. Move the pointer to the first data point in the
mA (or VDC)
column and press [YES]. The units for this column will depend on
the units you selected in Activating and Changing Probes on
page 2-9.
8. Enter the voltage or current value and press [YES].
9. Move the pointer to the
Units column and press [YES].
10. Enter the corresponding scale value and press [YES].
11. Use steps similar to 4 through 10 to enter the remaining data
point values.
Repeat steps 1 through 9 to enter auxiliary input calibration data for
the remaining channels.
To enter calibration data for moisture, temperature, and pressure,
refer to the preceding sections. To exit, press the DONE menu key
until
Main Menu appears on the message line. Then refer to Entering
High and Low Reference Values on page 2-21.