Basic Programming 2-25
April 2004
Entering Delta F Oxygen
Cell Reference Values
(cont.)
3. Move the pointer to the Zero line for the high reference for the
desired channel and press [YES].
4. Enter the zero value for the high reference for that channel and
press [YES]. The pointer automatically proceeds to the
Span line
for the high reference for the same channel. Press [YES] again.
5. Enter the span value for the high reference for that channel and
press [YES].
6. Move the pointer to the
Zero line for the low reference for the
desired channel and press [YES].
7. Enter the zero value for the low reference for that channel and
press [YES].
8. Move the pointer to the Span line for the low reference for the
desired channel and press [YES].
9. Enter the span value for the low reference for that channel and
press [YES].
Repeat steps 3 through 9 to enter the high and low reference values
for the remaining channels.
To enter reference values for moisture and pressure inputs, refer to
Entering Moisture Reference Values on page 2-22 and Entering
Pressure Reference Values on page 2-26. To exit, press the DONE
menu key until
Main Menu appears on the message line. Then refer
to Displaying Measurements on page 2-28. When you exit, the
Series 1 automatically calibrates each channel.