April 2004
3-16 Advanced Programming
Using Computer
Enhanced Response
(cont.)
Figure 3-9: Enhanced Response Menu
4. Use the arrow keys to scroll to the desired channel.
5. Press the ON/OFF menu key to turn enhanced response on (yes) or
off (no).
6. Repeat steps 4 and 5 for each desired channel.
7. To exit, press the DONE menu key until Main Menu appears on
the message line.
8. Expose the sensor to dry gas.
IMPORTANT: You must enable enhanced response before you
expose the sensor to the dry gas. The meter must see
all of the dry down in order to make accurate
calculations.
When activating enhanced response for a channel, a reverse video
“E” symbol appears in the top-right-hand corner of the corresponding
box when displaying data in the matrix format. When the meter
determines the final value, the reverse video “E” changes to a regular
“E”. The “E symbol does not appear in the graph format.
HELP ON/OFF PAGE DONE
ENHANCED RESPONSE TABLE
Ch Enhanced
1
3
5
Enhanced Response Menu
2
4
6