2.3.8 Layout Constraints
NOTICE
The X-Ray tube Chiller elevation cannot exceed 3 meters (9.84 feet) in height.
NOTICE
Minimum distance between Digital Detector and UPS is 3 m (118 inches) (Fluoro
UPS option).
NOTICE
The distance of the Detector chiller cannot exceed 3 m (1 floor) in height below
the detector.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements 103