Chapter 5 Electrical Requirements
1 Power Requirements
1.1 System Electrical Characteristics
NOTICE
It is the customer’s responsibility to ensure that electrical installation is compliant to local
regulations. To avoid risk of electric shock, this equipment must only be connected to a supply
mains with Protective Earth.
Power supply and ground cables shall be dedicated to the system. They must not be used to
supply other systems.
Power supply and ground cables shall be kept separated from room system cables.
Power supply and ground cables must be connected to the same distribution panel. They must
run near one to the other.
Power supply and ground cables provided by the customer shall be compliant with local
regulations (e.g. UL, NFPA 70, CSA, IEC, CCC).
1.1.1 Core system
Table 5-1:
Nominal voltage Frequency
Power consumption Type of power input
Nominal Peak
Without Fluoro
UPS option
With Fluoro UPS
option
380 V ± 10%
50 Hz or 60 Hz (± 3 Hz)
60 kVA 150 kVA 3~ 3N~
400 V ± 10%
415 V ± 10%
480V ± 10% 60 Hz only (± 3 Hz)
The Hospital circuit breaker should fit the current protection of the system:
150A/ 480V, 3 Phases for UL
80A/ 380V, 3 Phases for CE
WARNING
THIS FLUORO UPS IS ONLY DESIGNED TO OPERATE IN A WYE-
CONFIGURED ELECTRICAL SYSTEM WITH A SOLIDLY GROUNDED
NEUTRAL.
FOR MORE DETAILS, REFER TO THE FLUORO UPS DOCUMENTATION.