EasyManua.ls Logo

GE Optima IGS 320 - 2286398 - on Grade, Floor Anchor; 2285052 - above Grade, Floor Anchor with Insert

GE Optima IGS 320
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.3.3.2.3 2286398 – On grade, Floor anchor
Table 2-14:
Item Name Part # Description Quan. Notes
Floor An‐
chor
46-302265P1 Floor Anchor Bolt; 5/8-10 x 6 in.
12 anchors for LC Positioner and
4 anchors for Table
16 pc On Grade anchor mounting
method hardware
Grout assy 2285055 - 10 kg Powdered Mortar Ardex
K-15
- RTV Silicon Rubber Adhesive
- 18 mm making tape adhesive
- Open cell foam
1 kit Used in constructing LC
Positioner grout dam
Dowel 2290937-2 Wood Dowel; 16 mm diameter 12 pc
Cable Con‐
duit
2285057 Cable conduit - sheet metal part 1 pc
Vinyl Trim 2296890 Gripping Range; 1.5 to 3 mm THK 1 m
Vinyl Trim 2296891 Vinyl trim with segmented metal
core 12 mm
1 m
3.3.3.2.4 2285052 – Above grade, Floor anchor with insert
Table 2-15:
Item Name Part # Description Quan. Notes
Floor An‐
chor
46-302265P1 Floor Anchor Bolt; 5/8-10 x 6 in.
drilling
12 anchors for LC Positioner and
4 anchors for Table
16 pc Above Grade anchor
mounting method hardware
Grout assy 2285055 - 10 kg Powdered Mortar Ardex
K-15
- RTV Silicon Rubber Adhesive
- 18 mm making tape adhesive
- Open cell foam
1 kit Used in constructing LC
Positioner grout dam
Dowel 2290937-2 Wood Dowel; 16 mm diameter 12 pc
Water Elec‐
tric Separa‐
tor
2268647 Manufactured part introduced in
Positioner key hole to fit 1 electri‐
cal conduit of 6”.
1 pc Electrical and water con‐
duits separated inch size.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
126 3 Room Structural Requirements