EasyManua.ls Logo

GE Optima IGS 320 - 2 Room Layouts

GE Optima IGS 320
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 Room Layouts
2.1 Room Dimension Requirements
WARNING
LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS
MANDATORY.
THE ELECTRONIC CABINETS (SYSTEM CONTROL CABINET, COOLIX 4100
CHILLER, DETECTOR CHILLER AND FLUORO UPS WHEN INSTALLED )
INCLUDE FANS THAT ARE CREATING AIR–CIRCULATION OF PULSED–AIR.
WHEN THIS PULSED AIR IS IN AN ENVIRONMENT THAT MAY CONTAIN
AIRBORNE PATHOGENS LIKE AN EXAM ROOM/CONTROL ROOM, THERE
IS A RISK OF TRANSMISSION OF THESE AIRBORNE PATHOGENS FROM
PATIENTS TO OTHER PATIENTS OR CLINICAL PERSONNEL (NOSOCOMIAL
DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE
INSTALLED IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM,
I.E., TECHNICAL ROOM.
For exam and control rooms, several configurations are recommended.
NOTE: Motion controls installed in remote location from the table shall be installed at a
location where all the positioner axis are visible by the operator.
Refer to Patient Room Layout and Control Room Layout to see recommended exam / control
rooms layouts.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements 85

Table of Contents

Related product manuals