3.3 Door Size Requirements
Minimum door sizes also apply to hallways and elevators. For additional details, refer to
Shipping Information.
3.3.1 Door Height
The minimum door height (to accommodate system positioner on its dolly and System Control
Cabinet) is 2.18 m (86 in).
3.3.2 Door Width
The minimum door width needed (to accommodate the System LC shipping dolly) is:
1.165 m (46 in) with protective side rail,
1.096 m (43 in) with one protective side rail removed on site.
NOTE: Door widths are based on a
straight–in
approach requiring a 2.44 m (96 in) wide
corridor. Calculations need to be made for accommodation of equipment through
narrower corridors.