Table 2-10: D distance
ANGIO CARDIO NEURO
Omega IV Compact NA 1395 mm (54.9 in) NA
Omega V Long 1278 mm (50.3 in) 1395 mm (54.9 in) 1395 mm (54,9 in)
3.3.3 LC Positioner and Omega Table Floor Preparation Kits (GEMS supplied)
Illustration 2-64:
All GE supplied vascular system floor preparation are contained in catalog. There are some
additional gantry/table mounting kits based on each mounting method: through bolts or floor
anchors.
Base plate assembly (mandatory) 2285048 – Refer to Table 2-11
If there is no integrated base plate plan (2359194) delivered with S18101SG, please refer to
drawings 2359194 (Angio and Cardio configuration Illustration 2-65) Please order locally the
integrated base plate based on this plan.
Assembly and separation (Select kit) – Refer to tables, Section 3.3.3.1
Above grade, Floor anchors without insert – 2285051.
Through bolts without insert – 2285646.
On grade, Floor anchors – 2286398.
Above grade, Floor anchors and insert – 2285052.
Through bolts and insert – 2285053.
Floor plate / Base plate Assembly – 2285632.
Base Plate Assembly Kit - 5570895