Installation
2030047-001A Patient Data Module 3-3
4. Press the battery door closed until it seals the battery compartment.
WARNING
PHYSICAL INJURY—Make sure the battery is completely inserted
and that the battery door is securely sealed.
Falling batteries could seriously or fatally injure neonatal or other
vulnerable patients.
5. Press the Power button on the Patient Data Module.
6. Verify that the Power indicator illuminates amber while the Patient Data Module
boots up, then illuminates green.
Software
The Patient Data Module leaves the factory with software installed. If you need to
install the software, contact your GE representative.
Mounting options
WARNING
For safety reasons, all connectors for patient cables and sensor leads
are designed to prevent inadvertent disconnection, should someone
pull on the leads. Do not route cables in a way that they may present
a stumbling hazard.
WARNING
PHYSICAL INJURY—Do not hang articles on the IV pole that are
not related to the Patient Data Module’s use.
Do not place the Patient Data Module more than 147 cm (58 in)
from the floor when mounting on an IV pole with a base not less
than 58 cm (23 in) in diameter.
WARNING
To reduce the ingress of water into the equipment, do not mount the
Patient Data Module in a vertical position with the patient cables
facing up or down.