Field Replaceable Units
2030047-001A Patient Data Module 7-13
NBP assembly with manifold hose
1. Remove the top housing, main board, NBP hose, coupling and elbow per above
steps.
2. Remove 4 screws from the battery cover.
467A
3. Grasp and slide the battery cover out to the rear.
4. Remove 1 screw under the pump tube connection.
466A
5. Lift off the manifold with pump.
6. Remove the interface gasket.
Front panel board connector Main board flex pump connector
Battery cover