Loading...
7-6 Patient Data Module 2030047-001A
Field Replaceable Units
Labels
Apply labels as shown.
404B
464A
Battery door and tray
1. Remove 1 screw that holds the battery door to the housing and remove the door
and washer.
462A
2. Remove the battery if one is in the tray.
3. Use a flathead screw driver to lift the catch tab on the tray as you pull it out of the
housing.