Appendix A–Technical Specifications
2030047-001A Patient Data Module A-5
Respiration
Temperature
Respiration range limit: 1 to 200 breaths/minute
Respiration rate accuracy 0 to 120 breaths/minute ± 1
121 to 200 breaths/minute ± 3
Impedance range: 100 to 1500 Ohms at 52.6 KHz
Detection sensitivity range: 0.4 to 10 Ohms impedance variation
Respiration rate alarm
range:
1 to 200 breaths/minute
No breath alarm range: 3 to 30 seconds
Number of channels: Up to 2 (with Y-adapter cable)
Input specifications
Probe type: Series 400 or 700 (determined by input cable)
Temperature range: 0° C to 45° C (32° F to 113° F)
Resolution: ±0.01° C (internal)
Output specifications
Parameters displayed: T1, T2
Error (independent of source):
±0.1° C with series 400 probes
±0.3° C with series 700 probes
Alarms: User-selectable upper and lower limits