Field Replaceable Units
2030047-001A Patient Data Module 7-5
855A
4. Press the battery door closed until it seals the battery compartment.
WARNING
PHYSICAL INJURY—Make sure the battery is completely inserted
and that the battery door is securely sealed.
Falling batteries could seriously or fatally injure neonatal or other
vulnerable patients.
5. Press the Power button on the Patient Data Module.
6. Verify that the Power LED illuminates amber while the Patient Data Module
boots up, then illuminates green.
404B
Power LED