EasyManuals Logo

GE Phasor CV User Manual

GE Phasor CV
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
3. Operating in Phased Array Mode
Page 42 PHASOR XS Operating Manual
Carefully calibrate the instrument for the probe /
material combination prior to beginning the TCG
recording process.
Set the MAT THICKNESS and LEG values so that
all reference reflectors will fall within the displayed
sector. These settings can not be adjusted while
recording TCG points.
TCG reference points (up to 15) must be recorded
by obtaining an A-Scan echo of a representative
reflector or manually entered using the TCG EDIT
Submenu. The reference points allow the instru-
ment to calculate and compensate for the effect of
material attenuation on reflector-amplitude height.
The dynamic range of the TCG function is 40 dB.
Maximum curve slope is 6 dB per microsecond.
Successive data points do not have to decrease in
amplitude. That is, the TCG curve does not have to
have a constantly descending slope.
3.6.1 Recording the TCG Reference
Points
TCG reference points are typically taken from a standard
with equally sized reflectors (holes) located at various
material depths. The primary echo from each of these
points (for up to a total of 15 echoes) are recorded. When
TCG is active, the instrument compensates for different
material thickness and different detecting-beam angles
by applying a varying gain level to echoes at material
depths other than the baseline depth. Only one set of
TCG reference points can be stored at a time. To record
TCG reference points:
Step 1: A-Gate is used to record the reference echoes.
Set GATE START and WIDTH to evaluate only the se-
lected reference echo. Set THRESHOLD to 5% in order
to include any signal height above 5% FSH.
Step 2: Access the NRM/TCG Submenu (located in the
UT Menu) by pressing
.
Step 3: Start the recording process by pressing (two
times)
next to RECORD so that POINT 1 appears.
Couple the phased array probe to the test standard so that
at least one “step” segment on the display’s left side (see
Figure 3-9) indicates the reflector’s presence. As seen in
Figure 3-9, display segments move to the right from their
baseline position by an amount that’s proportionate to their
maximum amplitude. Note that changing Gain setting or
moving the A-Gate THRESHOLD will alter the amplitude
at which this reflector is detected in A Gate.
NOTE: During the recording process, the gain applied
to one or more beam group segments can be reduced
to eliminate the effect of an undesired reflector (such
as a geometric feature that is detected by these spe-
cific segments) on the recording of a TCG point. See
Section 3.6.2 to select a region of interest during the
TCG-point recording process.
NOTE: The cyclic gain feature (
Section 3.6.3) adds an
incremental amount of gain to each beam group.
FIGURE 3-8—The Amplitude Cycle Diagram indicates the TCG-reference amplitude recorded at each beam angle.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Phasor CV and is the answer not in the manual?

GE Phasor CV Specifications

General IconGeneral
BrandGE
ModelPhasor CV
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals