EasyManua.ls Logo

GE Phasor CV

GE Phasor CV
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. Conventional Operation: Menu System, Keypad, and Displays
Page 60 PHASOR XS Operating Manual
note that the range can vary from 0.040 to 300 inches.
Step 4: The display’s horizontal range will remain as
set.
4.8.2 Setting the Display Delay
The display delay function shifts the displayed A-Scan to
the left or right. This function is used to set the instrument’s
viewing window. To set the display delay
Step 1: Activate the Home Menu by pressing
. Func-
tions will appear down the left side of the display sreen.
Step 2: Press
next to the selection titled DISPLAY
DELAY.
Step 3: To change the display delay turn the function knob.
You’ll note that the displayed echoes shift to the left or
right.
4.9 Calibrating the Instrument
4.9.1 Pre-calibration Check List
To improve the accuracy and quality of your calibration, be
sure that the following conditions are met before launching
the calibration function:
• Probe installed
DUAL (RECEIVER) setting must match probe
Set the material type (
Section 2.4)
Recommended that DISPLAY DELAY be set to 0
PRF set to AUTO.
• TCG—Turned OFF
Recommended that REJECT be set to 0.
4.9.2 Using AUTOCAL to Calibrate the
Instrument
(Read the following while reviewing Figure 4-12)
Step 1: From the SETUP Menu, activate the AUTOCAL
Submenu by pressing
below it. Four functions will
appear down the left side of the display screen.
Step 2: Press
next to the selection titled S-REF1 and
turn the function knob until the value of S-REF1 matches
the thickness of the thinner calibration standard.
Step 3: Press
next to the selection titled S-REF2 and
turn the function knob until the value of S-REF2 matches
the thickness of the thicker calibration standard.
Step 4: Apply couplant and couple the probe to the thin-
ner calibration standard. Press
next to the selection
titled A START. Turn the function knob (this will shift the
starting point of the A-Gate) until the A-Gate lies over the
echo corresponding to the thinner standard’s thickness
(
Figure 4-12).
Step 5: Press
next to the selection titled RECORD.
The value in the function box will change from “OFF” to
“S-REF1?”. While maintaining the signal in the A-Gate,
NOTE: The instrument’s AUTOCAL Function applies
only to conventional operation, not to phased-array
operation.
FIGURE 4-11—Effect of A-Scan Range Adjustment

Table of Contents

Related product manuals