MAINTENANCE
GE Medical Systems
SENOGRAPHE 800T
REV 6 om 2135575–100
3–6
ÁÁÁ
ÁÁÁ
Message #
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Message on Control Console
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Explanation / Meaning
of Message
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Corrective Actions
ÁÁÁ
ÁÁÁ
I 07
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
SELECTED FILTER NOT
ALLOWED: xx
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
The Filter selected by the oper-
ator is not allowed
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Information message only, because the
Senographe Automatically selects the
correct filter.
ÁÁÁ
ÁÁÁ
I 12
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
MAXIMUM mAs = xx
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Maximum available mAs are
equal to xxx
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Information message only. (If mAs
exceeded, cell aborts exposure at 2 or 3
mAs if entire exposure cannot be
made).
ÁÁÁ
ÁÁÁ
ÁÁÁ
I 20
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
AEC FAILURE
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
The AEC is faulty
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service. You can
still work in Manual mode.
ÁÁÁ
ÁÁÁ
I 21
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
HEATING FAILURE TRACK 1
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Large focal spot Track is out of
use
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service. It is still
possible to use the small focal spot.
ÁÁÁ
ÁÁÁ
ÁÁÁ
I 22
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
HEATING FAILURE TRACK 2
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Small focal spot Track is out of
use
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service. It is still
possible to use the large focal spot.
ÁÁÁ
ÁÁÁ
I 23
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
CONSOLE COMMUNICATION
FAIL
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
generator to Control Console
connection is faulty.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service.
ÁÁÁ
I 24
ÁÁÁÁÁÁÁÁ
LINE POWER INTERRUPTION
ÁÁÁÁÁÁÁÁ
Micro power cut occurred on
line supply.
ÁÁÁÁÁÁÁÁÁ
Press RESET to suppress message.
I 25
POWER SUPPLY DEFECT
Call Service
ÁÁÁ
I 27
ÁÁÁÁÁÁÁÁ
CPU BATTERY ERROR
ÁÁÁÁÁÁÁÁ
CPU back up battery is dis-
charged
ÁÁÁÁÁÁÁÁÁ
Contact GEMS field service. DO
NOT SWITCH THE EQUIPMENT
OFF
ÁÁÁ
ÁÁÁ
ÁÁÁ
ÁÁÁ
ÁÁÁ
ÁÁÁ
ÁÁÁ
I 28
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
MAX. WAIT FOR SAME EXPO-
SURE: xx min.
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
It is necessary to wait the dis-
played amount of time to be
able to repeat the same expo-
sure.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Wait the required amount of time to
repeat exposure. If an exposure is
attempted before this time, an abort
MAY occur (we have built in a safety
margin to ensure proper operation). It
also means that you can ignore the
message if a different, less dense,
breast is to be imaged next or if differ-
ent technical factors are selected.
ÁÁÁ
ÁÁÁ
ÁÁÁ
I 29
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
EXCESSIVE mAs FOR THIS
MODE
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
The amount of mAs necessary
to repeat the same exposure
may be too large in the mode
selected by the user.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch to a different mode or to a dif-
ferent filter if in AEC mode.