MAINTENANCE
GE Medical Systems
SENOGRAPHE 800T
REV 6 om 2135575–100
3–19
By increments of 1 cm to maximum
= 20 cm.
By decrements of 1 cm to minimum
= 1 cm.
To confirm.
Note:
The automatic decompression height that is being programmed represents the height
from the last position of the compression paddle.
Example:
If the compressed breast thickness was 4.6 cm before decompression and a 20cm
automatic decompression height has been programmed, the paddle will go back up to
24.6cm.
If the compressed breast thickness was 8.3 cm before decompression and a 20cm
automatic decompression height has been programmed, the paddle will go back up to
28.3 cm.
Etc, ...
4–4–3 Confirm & Return to Application Menu
To confirm the function selected and return to the APPLICATION Menu (daily use), press
repeatedly until focal spot, focal track, filter, etc” return.