SPECIFICATIONS
GE Medical Systems
SENOGRAPHE 800T
REV 6 om 2135575–100
4–2
SECTION 2
TECHNICAL SPECIFICATIONS
2–1 Electrical Specification
2–1–1 Line Power Supply
Input voltage, single-phase: 200/208/220/230/240V (± 10%).
Line frequency: 50 or 60 Hz (± 1 Hz).
2–1–2 Maximum Line Current
Line power: 6 kVA during exposures
Maximum line current:
– 200V: 30A
– 240 V: 25A
2–1–3 Permissible Line Resistance
The maximum line impedance authorized are the following, depending on the mains nominal
voltage:
200V ––––> 0.33 Ohms
208V ––––> 0.36 Ohms
220V ––––> 0.40 Ohms
230V ––––> 0.44 Ohms
240V ––––> 0.48 Ohms
2–1–4 Generator Output
22 through 35 kV,
20 through 100 mA.
2–1–5 Anode Current
I max large focus = 100 mA at 30 kV,
I max small focus = 30 mA at 30 kV.
2–1–6 Generator Power
3 kW maximum.
2–1–7 Duty Cycle
The generator is always limited by the tube:
The generator can supply an average power 100 W to the tube.
This example represents one 30 kV/200mAs exposure each minute.