MAINTENANCE
GE Medical Systems
SENOGRAPHE 800T
REV 6 om 2135575–100
3–5
SECTION 3
MESSAGES
Three types of messages may be displayed for the user on the gantry or on the Control
Console:
I xx: Information message: fault message which does not inhibit the exposure.
E yy: Error message: fault inhibits the exposure.
S zz: Stop message: exposure is stopped due to incorrect use of equipment.
A message displayed on the Control Console has 40 characters. The first three characters are
identification characters. The next 28 characters comprise the message. The seven remaining
characters are also identification characters.
Example: “I69 MAGNIFICATION FAILURE 172/088”
When calling the Field Service of GEMS, it is essential to note each message in its entirety
Listed below by type are all the messages which can appear on the Control Console, together
with a detailed description and the necessary corresponding corrective action.