MAINTENANCE
GE Medical Systems
SENOGRAPHE 800T
REV 6 om 2135575–100
3–30
4. To remove the bulb from its housing, hold it by its white part (1) and raise the lever (2).
Discard the used bulb.
1
2
5. Remove the packing of the new bulb.
ONLY HOLD THE BULB BY ITS WHITE PART. DO NOT TOUCH THE
GLASS PART WITH BARE FINGERS OR THE BULB MAY BE
PERMANENTLY DAMAGED.
IF YOU ACCIDENTLY TOUCH THE GLASS PART OF THE BULB
WITH YOUR SKIN CLEAN THE BULB CAREFULLY WITH A SOFT
CLOTH CAMPENED WITH ALCOHOL.
TO AVOID ANY INCIDENTS YOU ARE ADVISED TO WEAR
COTTON OR LATEX GLOVES WHEN REPLACING THE BULB, OR
TO USE THE BULB PACKING WHEN HANDLING THE BULB.
6. Align the two different–sized pins of the new bulb with its socked and push the bulb
upwards.
7. Close the bulb protection cover by pulling it down and pivoting it.
8. Refit the collimator cover (see illustration in point 2 on page 3–28).
FOR ADEQUATE PROTECTION, AFTER REPLACING THE LAMP, AND
BEFORE SWITCHING THE SYSTEM ON, REPLACE THE COLLIMATOR
COVER.
9. Switch on the Senographe 800T.
WARNING
WARNING