ElectromagneticCompatibility(EMC)
Guidanceandmanufacturer’sdeclaration–electromagneticimmunity
ImmunityTest
Compliance
Testlevel
CompliancelevelElectromagnetic
Environment–
Guidance
Electrostatic
discharge(ESD)
IEC60601-1-2
EN61000-4-2
±2/4/6kVindirect
±2/4/6kVdirect
±2/4/8kVair
±2/4/6kVindirect
±2/4kVdirect
±2/4/8kVair
Floorsshouldbe
wood,concreteor
ceramictile.Ifoors
arecoveredwith
syntheticmaterial,
therelativehumidity
shouldbeatleast
30%.
Electricalfast
transient/burst(EFT)
IEC60601-1-2
EN61000-4-4
±2kVforpower
supplylines
±1kVfor
input/outputlines
±2kVforpower
supplylines
±1kVfor
input/outputlines
Mainspower
shouldbethatof
atypicalcommercial
orhospital
environment.
FastTransientSurge
(FTS)
IEC60601-1-2
EN/IEC61000-4-5
±500V/1kV
differentialmode
±2kVcommonmode
±500V/1kV
differentialmode
±2kVcommonmode
Mainspower
shouldbethatof
atypicalcommercial
orhospital
environment.
Voltagedips,short
interruptionsand
voltagevariationson
powersupplyinput
lines
IEC60601-1-2
EN/EC61000-4-11
<5%U
t
(>95%dipin
Ut)for0.5cycles
<40%U
t
(>60%dipin
Ut)for5cycles
<70%U
t
(>30%dipin
Ut)for25cycles
<5%U
t
(>95%dipin
Ut)for5s
<5%U
t
(>95%dipin
Ut)for0.5cycles
<40%U
t
(>60%dipin
Ut)for5cycles
<70%U
t
(>30%dipin
Ut)for25cycles
<5%U
t
(>95%dipin
Ut)for5s
Mainspower
shouldbethatof
atypicalcommercial
orhospital
environment.If
theuserrequires
continuedoperation
duringpowermains
interruptions,itis
recommendedthat
powerbesupplied
fromanapplicably
rateduninterruptible
powersupplyora
battery.
Powerfrequency
(50/60Hz)magnetic
eld
IEC60601-1-2
EN/IEC61000-4-8
3A/m3A/m
Powerfrequency
magneticelds
shouldbeatlevels
characteristicsofa
typicallocationina
typicalcommercial
orhospital
environment.
NOTE:U
t
isthea.c.mainsvoltagepriortoapplicationofthetestlevel.
ElectromagneticImmunity
TheT2100Treadmillisintendedforuseintheelectromagneticenvironmentspecied
below.ThecustomeroruseroftheT2100Treadmillshouldassurethatitisusedin
suchanenvironment.