TechnicalSpecications
Item
Specication
FloorSpaceRequired
81x390cm(32x154in)
390cm(153.5in)=205cm(80.7in)fortreadmill+185
cm(72.8in)ofclearance
WalkingSurfaceHeight26cm(10.2in)(ProductCodeSK2)
14.0cm(5.5in)(ProductCodeSM9)
HandrailHeightAboveWalking
Surface
Front:101.6cm(40in)
Side:Maximum88.9cm(35in)atthefrontdescending
ata5°angleto81.3cm(32in)
CeilingHeight(minimum)
270cm(106in)
Power/EnvironmentalSpecications
Item
Specication
PowerRequirements
200to240VAC,5060Hz,single-phase,20
amperes,NEMA6-20Rwallsocket(U.S.only,or
applicableinternationalconnection;dedicatedcircuit
recommended)
PowerConsumption
1540watts(5254BTU),16amperes
Safety
Item
Specication
Certication
UL/cULclassied
TypeofProtectionAgainstElectrical
Shock
ClassI
DegreeofProtectionAgainst
ElectricalShock
TypeBappliedpart
DegreeofProtectionAgainstIngress
ofLiquids
IPx0
Ordinaryequipment(enclosedequipmentwithout
protectionagainstingressofwater).
HandlingofDisposableSuppliesand
OtherConsumables
Useonlypartsandaccessoriesmanufacturedor
recommendedbyGEMedicalSystemInformation
Technologies.
Followmanufacturer’sinstructionsforusefor
disposable/consumableproduct.
Followlocalenvironmentalguidelinesconcerningthe
disposalofhazardousmaterials(forexample,lead
acidbatteries).
PatientLeakageCurrent
100/10μA(AC/DC)innormalcondition
500/50μA(AC/DC)insinglefaultcondition