Introduction/Safety
1-26 Venue 50 – Basic User Manual
Direction 5446729-100 English
Rev. 7
Declaration of Emissions
This system is suitable for use in the following environments.
The operator must assure that the system is used according to
the specified guidance and only in the electromagnetic
environment listed below.
Table 1-3: Declaration of Emissions
Guidance and manufacturer’s declaration - electromagnetic emissions
The system is intended for use in the electromagnetic environment specified below. The
operator of the system should assure that it is used in such an environment.
Emission Type Compliance Electromagnetic Environment
RF Emissions
CISPR 11
Group 1 This system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A This system is suitable for use in all establishments, other than
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes, provided the following warning is
heeded:
WARNING: This system is intended for use by healthcare
professionals only. This system may cause radio interference or
may disrupt the operation of nearby equipment. It may be
necessary to take mitigation measures, such as re-orienting or
relocating the system or shielding the location.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
Fluctuations/Flicker
Emissions
IEC 61000-3-3
Complies