Preparing the System for Use
2-4 Venue 50Basic User Manual
Direction 5446729-100 English
Rev. 7
Environmental Requirements
The system should be operated, stored, or transported within
the parameters outlined below. Either its operational
environment must be constantly maintained or the unit must be
turned off.
Acclimation Time
After being transported, the unit requires one hour for each 2.5
degree increment when its temperature is below 10°C or above
40°C.
Table 2-1: System Environmental Requirements
Operational Storage Transport
Temperature 3 - 40°C
37 - 104°F
-5 - 50°C
23 - 122°F
-5 - 50°C
23 - 122°F
Humidity 30 - 80% non-condensing 10 - 90% non-condensing 10 - 90% non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
Table 2-2: System Acclimation Time Chart
°C 60 55 50 45 40 35 30 25 20 15 10
°F 140 131 122 113 104 95 86 77 68 59 50
hours 86420000000
°C 5 0 -5 -10 -15 -20 -25 -30 -35 -40
°F 41 32 23 14 5 -4 -13 -22 -31 -40
hours 2468101214161820