Introduction/Safety
1-32 Venue 50Basic User Manual
Direction 5446729-100 English
Rev. 7
Best practices while scanning
Acoustic Output Default Levels
In order to assure that an exam does not start at a high output
level, the Venue 50 initiates scanning at a reduced default
output level. This reduced level preset depends upon the exam
application and probe selected.
To modify acoustic output, go to Utility->Image->User Define,
then adjust the acoustic output level in AO select box.
HINTS
Raise the Acoustic Output only after attempting image
optimization with controls that have no effect on Acoustic
Output, such as Gain and TGC.
WARNING
Be sure to have read and understood control explanations for
each mode used before attempting to adjust the Acoustic
Output control or any control that can effect Acoustic Output.
A
coust
i
c
Output
Hazard
Use the minimum necessary acoustic output to get the best
diagnostic image or measurement during an examination.
Begin the exam with the probe that provides an optimum focal
depth and penetration.