Introduction/Safety
1-34 Venue 50Basic User Manual
Direction 5446729-100 English
Rev. 7
“Consult accompanying documents” is
intended to alert the operator to refer
to the operator manual or other
instructions when complete
information cannot be provided on the
label.
Various
“CAUTION” - Dangerous voltage (the
lightning flash with arrowhead) is used
to indicate electric shock hazards.
Various
“ON” indicates the power on position
of the power switch.
CAUTION: This Power Switch DOES
NOT ISOLATE Mains Supply.
See the Console Overview
section for location
information.
“Protective Earth” indicates the
protective earth (grounding) terminal.
• Inside of AC adapter with
docking station
• Docking cart
“Equipotentiality” indicates the
terminal to be used for connecting
equipotential conductors when
interconnecting (grounding) with other
equipment.
Connection of additional protective
earth conductors or potential
equalization conductors is not
necessary in most cases and is only
recommended for situations involving
multiple equipment in a high-risk
patient environment to provide
assurance that all equipment is at the
same potential and operates within
acceptable leakage current limits. An
example of a high-risk patient would
be a special procedure where the
patient has an accessible conductive
path to the heart such as exposed
cardiac pacing leads. IEC60417-5021
Docking Cart
NRTL Listing and Certification Mark is
used to designate conformance to
nationally recognized product safety
standards. The Mark bears the name
and/or logo of the testing laboratory,
product category, safety standard to
which conformity is assessed and a
control number.
Back of Venue 50
Table 1-5: Label Icons
Label/Icon Purpose/Meaning Location