IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
Chapter 5 - Components and functions (theory) 5 - 9
5-3-3 VIVID E9 / VIVID E7 general description
VIVID E9 / VIVID E7 is a digital beamforming system. Signal flow travels from the Probe Connector
Panel to the Front End Electronics, then to the Back End Processor, and finally, the results are displayed
on the monitor.
System configuration is stored on the hard drive, inside the Back End Processor (BEP), and all
necessary software is loaded from the hard drive on power up.
A Physio module, the Patient I/O, is incorporated in the Back End Processor (BEP) to provide ECG
signals to synchronize cardiac ultrasound image acquisition. Other analog signals, from devices such
as treadmills (e.g. ECG and phono), may also be processed by the Patient I/O.
5-3-4 VIVID E9 / VIVID E7 block diagram
NOTE: Depending on VIVID E9 / VIVID E7 model and installed options, the number of cards and the
part numbers on the cards will vary.
Figure 5-4 VIVID E9 / VIVID E7 block diagram