IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
3 - 38 Section 3-8 - Connectivity setup
Section 3-8
Connectivity setup
NOTE: If connected to a stand-alone network (Peer-to-Peer network with a VIVID E9 / VIVID E7
scanner, an EchoPAC PC workstation and an optional network printer), you should use default
delivery settings.
3-8-1 Introduction
To be able to use the network functions when connected to a hospital network, the scanner must have
a proper network address.
• Before you can set up the scanner, you need to collect some information.
•The "Worksheet for DICOM Network Information" on page 2-14 can be used for gathering this
information.
• Typical source for this information is the network administrator.
3-8-2 Compatibility
VIVID E9 / VIVID E7 can communicate with:
• EchoPAC PC
• Image Vault
• Other units via DICOM