IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
5 - 76 Section 5-11 - Input and Output (I/O) modules
Section 5-11
Input and Output (I/O) modules
5-11-1 Purpose of this section
This section describes the input/output modules on VIVID E9 / VIVID E7.
5-11-2 Patient I/O (Physio)
5-11-2-1 General description
NOTE: The ECG functionality of the Patient I/O module is not intended for patient monitoring nor to
support alarm functionality. This input is intended as a tool for easier synchronization of images
and cineloop control during ultrasound examinations.
The Patient I/O panel is located on the front of VIVID E9 / VIVID E7.
The Patient IO contains the electronics for:
• Analog inputs - AUX (Pressure/Pulse)
• Phono
• ECG/Respiration
Figure 5-49 Patient I/O Panel
1 - PHONO
2 - ECG
3 - AUX (PRESSURE/PULSE)