GE
D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
Chapter 5 - Components and functions (theory) 5 - 35
5-8-4-2 Location in the Unit
The Relay board is located in the Front End Rack on the end nearest to the front of the scanner.
5-8-4-3 Input DC Voltages:
+15VDC
•+6VDC
•-6VDC
-15VDC
+100V PMXVPP (Voltage for Probe MUX)
-100V PMXVNN (Voltage for Probe MUX)
Voltages provided by regulators on the board:
+12 VDC
•+5 VDC
+3.3 VDC
-3.3 VDC
•-5 VDC
and
LVDC (3.3 V/2.5A) for RT3D probe
5-8-4-4 Input Signals
Transmit period:
XMIT Pulses via the XD_BUS
Receive period:
Echo signals from the selected probe
5-8-4-5 Output Signals
Transmit period:
XMIT Pulses to the selected probe
Receive period:
Echo signals via XD_BUS to GRX cards