EasyManua.ls Logo

GE Vivid E7

GE Vivid E7
736 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
7 - 20 Section 7-9 - Troubleshooting
7-9-9 System Temperature Too High
When the temperature inside the VIVID E9 / VIVID E7 increase, the fan speed will increase to cool down
the system. If the air filters become too dusty, a higher fan speed is required to keep the air stream at
the needed level. When the fan speed increase, the fan noise will also increase.
If the cooling air stream is insufficient to stabilize the interior temperature within the operating
margin, the system will stop operating.
- Solution: Clean or replace the filters.
If a fan is worn out, it may be noisy, or stop working.
- Solution: Replace the fan(s).