IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
8 - 150 Section 8-6 - Top Console Parts Replacement Overview
8-6-31 Handle, Left Top / Handle Right Top, replacement
8-6-31-1 Preparations
1.) Power down the VIVID E9 / VIVID E7 and disconnect the Mains Power Cable.
2.) Remove the buttons below the Touch Panel.
3.) Remove the Upper Operator Panel.
4.) Remove the Lower Operator Panel.
Figure 8-132 Handle, Left, Top