2 - 2 Section 2-2 - General console requirements
Section 2-2
General console requirements
2-2-1 Console environmental requirements
2-2-1-1 If the unit is very cold or hot
When unpacking the Ultrasound system, allow the temperature of the Ultrasound system to stabilize
before powering up. The following table describes guidelines for reaching operational temperatures
from storage or transport temperatures.
2-2-1-2 Environmental specifications for VIVID E9 / VIVID E7 scanners
Temperature Limits:
•Operation:
10 to 35 ºC (50 to 95 ºF)
Storage and Transport:
-20 to 60 ºC (-4 to 140 ºF)
Humidity Limits:
•Operation:
30 - 60% rH non-condensing
Storage and Transport:
30 - 95% rH non-condensing
Air Pressure Limits:
•Operation:
700-1060 hPa
Storage and Transport:
700-1060 hPa
Heat Dissipation:
3800 BTU/h
CAUTION
IF THE UNIT IS VERY COLD OR HOT, DO NOT TURN ON ITS POWER UNTIL IT HAS HAD A
CHANCE TO ACCLIMATE TO ITS OPERATING ENVIRONMENT.
Table 2-1 VIVID E9 / VIVID E7 Acclimate Time
o
C
-40-35-30-25-20-15-10-5 0 5 10152025303540455055606570
o
F
-40 -31 -22 -13 -4 5 14 23 32 41 50 59 68 77 86 95 104 113 122 131 140 149 158
Hrs 2018161412108642000000024681012