GE
D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
Chapter 2 - Site preparations 2 - 5
2-2-2-2 Electrical requirements for VIVID E9 / VIVID E7
In the table below, the electrical specifications for VIVID E9 and VIVID E7 includes monitor and on board
peripherals.
The current drain will vary depending on the mains voltage.
At 230 VAC the current may be up to 5 A.
At 100 VAC the current may be up to 12 A.
2-2-2-3 Site circuit breaker
It is recommended that the branch circuit breaker for the unit be readily accessible.
2-2-2-4 Site power outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other outlets
adequate for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and
national electrical codes.
2-2-2-5 Unit power plug
If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
Table 2-2 Electrical specifications for VIVID E9 / VIVID E7
PART NUMBER DESCRIPTION VOLTAGE TOLERANCES
POWER
CONSUMPTION FREQUENCY
XDclear (S/W v113.x)
GB000070
Vivid E9 4D Expert Option 17 inch LCD
100-230 VAC ±10% 1100 W 50/60 Hz
GB000080
Vivid E9 4D Expert Option 19 inch LCD
GB000075
Vivid E9 2D 17 inch LCD
GB000085
Vivid E9 2D 19 inch LCD
GB000090
Vivid E9 Pro configuration 17 inch LCD
GB000095
Vivid E9 Pro configuration 19 inch LCD
GB000099
Vivid E7 'XDclear' 'Pro configuration' 17'
GB000100
Vivid E7 'XDclear' 'Pro configuration' 19'
CAUTION
POWER OUTAGE MAY OCCUR. THE VIVID E9 / VIVID E7 SCANNER REQUIRES A DEDICATED
SINGLE BRANCH CIRCUIT. TO AVOID CIRCUIT OVERLOAD AND POSSIBLE LOSS OF
CRITICAL CARE EQUIPMENT, MAKE SURE YOU DO NOT HAVE ANY OTHER EQUIPMENT
OPERATING ON THE SAME CIRCUIT.