Input and Output (I/O) modules
Vivid E80/E90/E95Service Manual 5-45
GC091052 Rev. 3
BEP I/O Board
General description
The BEP I/O board is the interface between the BEP and
the rest of the Ultrasound system.
The BEP I/O board is plugged into a connector on the BEP’s
main board.
The DVI signals are received via cable from the Display
GPU.
Location
The I/O board is located inside the BEP with some connectors
available on the rear of the Ultrasound system and some
available inside the Ultrasound system.
Connectors on the I/O Rear Panel
2 x USB 2.0 (via HUB)
Audio Line Out L+R
DVI-out (external monitor)
Giga bit LAN
Connectors on the I/O Side Panel (inside the Ultrasound system)
Test connector (used by manufacturing)
8 x USB 2.0
• 2 x eSATAp
DVI-D out (System Monitor)
• Digital Audio out
Operating Panel Audio out

Table of Contents