Configuration
Vivid E80/E90/E95Service Manual 3-37
GC091052 Rev. 3
Software Options configuration
Software Option introduction
A Software Option Key, an alphanumeric text string, enables a
software option or a combination of software options.
The Software Option Key is specific to each Ultrasound system.
NOTE: More than one Software Option Key may be in use, depending
on the installed options.
To install a Software Option
Follow these steps to install the Software Option Key:
1. Press Config (F2) and log on as adm.
2. Select Admin (lower part of window).
3. Select the System Admin tab.