Casters and Brakes replacement
Vivid E80/E90/E95Service Manual 8-81
GC091052 Rev. 3
Install the Pedal Mechanism
Follow this procedure to install the Pedal Mechanism:
1. Position the Pedal Mechanism and install the four fixing
screws (Torque: 20.5 Nm).
2. Install the two bottom screws, including the plate.
3. Install the screw that ties the rod assembly to the pedal
mechanism.
4. Connect the rods by snapping on the ball joints.
5. Verify that the pedals work as intended:
a. Push down brake pedal (from “free” position), release it
and verify the pedal to return quickly up to half position.
b. Try to move the Ultrasound system back and forward;
and verify that the brake pedal stay in brake mode.
c. Push down direction lock pedal and verify the pedal to
return quickly.
d. Move Ultrasound system and verify Casters stay in
direction lock mode.
e. At push down, observe the “Latches” to come over pivot
T-Shape, and interfere by engaging.
6. Install the front cover.
7. Install both side covers.
8. Install the foot rest bumper.

Table of Contents

Other manuals for GE Vivid E95

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E95 and is the answer not in the manual?

GE Vivid E95 Specifications

General IconGeneral
TypeCardiovascular Ultrasound System
Transducers2D, 3D, 4D
Display22-inch high-resolution OLED monitor
PortabilityCart-based
Imaging Modes2D, 3D, 4D, Color Doppler
Advanced FeaturescSound Architecture
ConnectivityDICOM, USB, Ethernet
ApplicationsCardiology, Vascular, Abdominal

Related product manuals