Loading...
Maintenance task schedule
Vivid E80/E90/E95Service Manual 10-5
GC091052 Rev. 3
Inspect Cables and
Connectors
Clean Ultrasound system
Clean Monitor and Touch
Panel
Inspect Wheels, Casters,
brakes and Swivel Locks
Mobile Ultrasound system:
Check Daily
Check Operating Panel
Movement
Mobile Ultrasound system:
Check Daily
Ultrasound system Leakage
Current Checks
also after corrective
maintenance
Peripheral Leakage Current
Checks
also after corrective
maintenance
Surface Probe Leakage
Current Checks
also after corrective
maintenance
Endocavity Probe Leakage
Current Checks
also after corrective
maintenance
Transesophageal Probe
Leakage Current Checks
also after corrective
maintenance
Surgical Probe Leakage
Current Checks
also after corrective
maintenance
Functional Checks also after corrective
maintenance
Table 10-1: Customer Care Schedule (Cont'd)
Service at Indicated Time
Daily
Weekly
Monthly
Per Facilities
QA Program