70 8555-9201-000 / 0109
Mechanical Separation / GEA Westfalia Separator
4.25 Alarm #38 - Min. speed monitoring (option)
MIN. SPEED
MONITORING
during operation
Alarm#38 ENTER
Text display Possible causes Action
Actual speed lower than the permitted
speed.
• Check speed.
Speed setting for minimum speed too
high.
• Check maximum speed setting.
Frequency not stable.
•
Check frequency.
Drive (50 or 60 Hz) of separator does
not match frequency on site.
• Check separator configuration for
the frequency with the frequency on
site.
Speed pickup defective.
• Check speed pickup electrically;
replace if necessary.
•
Deselect function if required.
Note :
Alarm #38 only active when speed pickup is installed!