Mechanical Separation / GEA Westfalia Separator
4.27 Alarm #40 - No water seal
NO WATER SEAL
During TIMER 35
Separation “Stop”
Alarm#40 ENTER
Text display Possible causes Action
Clean oil discharge pressure (product
discharge pressure) too high, valve V7
closed too far.
Check clean oil discharge pressure,
product discharge pressure.
Filling water valve V2 doesn’t open.
Check function V2.
Filling water volume too low, e.g. other
consumers at the same time.
Check filling water volume.
Filling water valve V2 or filter soiled.
Check valve V2 and filter.
Water pressure for hydraulics too high
or closing water pulse T06 too long.
Check water pressure with open
valve.
Check T06.
Separating zone displacement:
Product temperature too low.
Check temperature and setpoint
TAL.
Wrong sensing liquid pump fitted.
Diameter too large.
Check fitted sensing liquid pump as
described in the separator manual.
Pressure transmitter PT2 bridged or
set to 0.
Check setpoint PT2.
Help text information on the control system
ENT E R
Step 1 Acknowledge fault
HEL P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
NO WATER SEAL
Please check:
Clean oil discharge
Function V2
->
Page 2
Filling water volume
Valve V2 and filter
Water pressure
Timer 06
->
Page 3
Setpoint TAL
Setpoint PT2
back with ENTER
F1