8555-9201-000 / 0109 71
Mechanical Separation / GEA Westfalia Separator
4.26 Alarm #39 - PT1 function (option)
PT1 function
In clean oil discharge
Separation “Stop”
Alarm#39 ENTER
Text display Possible causes Action
No pressure drop at PT1 during the
ejection.
Check switching function.
Pt1 is soiled, catches or is defective.
Clean and check PT1.
PT1 is set too low.
Check switching point PT1.
No ejection, ejection monitoring has
not responded.
Check bowl hydraulics and ejection
monitoring.
System pressure too high in clean oil
discharge (product discharge).
Check delivery head at clean oil
discharge (product discharge).
Deselect function if required.
Help text information on the control system
ENT E R
Step 1 Acknowledge fault
HEL P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
PT1 FUNCTION
Please check:
Switching function PT1
PT1 soiled
->
Page 2
Switching point PT1
Bowl hydraulics
Ejection monitoring
Delivery head at
->
Page 3
clean oil discharge
back with ENTER
F1