8555-9201-000 / 0109 73
Mechanical Separation / GEA Westfalia Separator
4.28 Alarm #41 - Min. separating time
MIN. SEPARATING TIME
< 10 mins
Separation “Stop”
Alarm#41 ENTER
Text display Possible causes Action
Pressure transmitter PT2 not con-
nected, soiled or defective.
Pressure transmitter PT2
Check switching function.
Alarm bridging time T11 too short.
Check T11.
Setpoint at pressure transmitter PT2
set too high.
Check setpoint at pressure transmit-
ter PT2.
Water or solid content in the bowl too
high.
Pre-dewater product and/or check
solid content.
Too much filling water.
Check filling water volume.
During HFO operating mode and pa-
rallel mode, threaded pins are fitted in
the separating disc.
Remove the threaded pins from the
separating disc.
Filling water valve V2 leaks.
Check filling water valve V2.
Pre-set valve in centripetal pump
chamber cover blocked.
Check pre-set valve.
Help text information on the control system
ENT E R
Step 1 Acknowledge fault
HEL P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
MIN. SEPARATING TIME
< 10 mins
Please check:
Pressure transmitter PT2
->
Page 2
Timer 11
Setpoint PT2
Water content
Valve V2
->
Page 3
During HFO operating mode
Threaded pins in
separating disc
back with ENTER