4.29 Alarm #42 - SMS sensor fault (option)
SMS SENSOR FAULT
During SMS CHECK
During TIMER 44
Alarm#42 ENTER
Text display Possible causes Action
PT2 no pressure signal.
Check setpoint PT2.
Filling water valve V2 no function.
Check function V2.
T44 too short.
Check T44.
Solids space blocked
Pressure sensor line to PT2.
Clean bowl, check bore in centri-
petal pump chamber cover.
Bowl not closed.
Check bowl hydraulics.
Filling water capacity during T44 is too
low.
Check water capacity (0.1/s)
Pressure switch PT2 defective.
Check pressure switch PT2.
Incorrect switching point at PT2.
Check switching point PT2.
Deselect function if required.
Help text information on the control system
ENT E R
Step 1 Acknowledge fault
HEL P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
SMS SENSOR FAULT
Please check:
Setpoint PT2
Function V2
->
Page 2
Timer 44
Sludge space
Bore in centripetal pump
chamber cover
->
Page 3
Bowl hydraulics
Function PT2
back with ENTER
F1