Program combination 2511022-150
Wrapped instruments, textiles, porous load
Parameter
1. Sterilizing temperatur
2. Sterilizing time
3. Postvacuum time
4. Postpuls steam
5. Postpuls air
°C
min
min
min
min
Range
134
3 - 7
0 - 90
0 - 90
0 - 90
Delivered
134
4
5
0
0
Actual
_ _ _ _ _ _
_ _ _ _ _ _
_ _ _ _ _ _
_ _ _ _ _ _
_ _ _ _ _ _
Heat sensitive material, rubber, plastic, porous load
Parameter
1. Sterilizing temperatur
2. Sterilizing time
3. Postvacuum time
4. Postpuls steam
5. Postpuls air
°C
min
min
min
min
Range
121
16 - 20
0 - 90
0 - 90
0 - 90
Delivered
121
16
5
0
0
Actual
_ _ _ _ _ _
_ _ _ _ _ _
_ _ _ _ _ _
_ _ _ _ _ _
_ _ _ _ _ _
Rapid process for single, open instrument
Password required!
Parameter
1. Sterilizing temperatur
2. Sterilizing time
3. Postvacuum time
°C
min
min
Range
134
3 - 90
0 - 90
Delivered
134
4
3
Actual
_ _ _ _ _ _
_ _ _ _ _ _
_ _ _ _ _ _
Bowie & Dick
Password required!
Parameter
1. Sterilizing temperatur
2. Sterilizing time
3. Postvacuum time
°C
min
min
Range
121 - 135
0 - 15
0 - 90
Delivered
134
1
3
Actual
_ _ _ _ _ _
_ _ _ _ _ _
_ _ _ _ _ _
Automatic leak rate test
Password required!
Parameter
1. Stabilizing time
2. Test time
min
min
Range
10 - 90
10
Delivered
10
10
Actual
_ _ _ _ _ _
_ _ _ _ _ _